The Series 2000 represents a complete line of Thermal Conductivity Detector (TCD) based analyzers which can be applied in a wide range of applications and industries. By using field proven filament-based and semiconductor based TC detectors, Teledyne is able to continuously monitor hydrogen and a variety of other gases of interest in either binary or multi-component sample gas streams.
- Three ranges plus cal range, field selectable application dependent, Auto Ranging
- Flush panel mount configuration
- Isolated 4-20 mADC and 0-1 VDC signal outputs and range ID contacts
- Programmable Auto Calibration Capability
- Five digit gas concentration LED display and separate back-lit 2 x 20 line alpha numeric vacuum fluorescent display for analyzer set-up / readout
- Nickel-plated brass cell-block with nickel alloy filaments and stainless steel end plates
- Self-diagnostics with Form C Failure Alarm Relay
- Two fully programmable concentration alarm set points and corresponding Form C + Failure Alarm Relay
- Linearizer: automatically linearizes the signal over the ranges and gases desired for analysis, and is programmed to provide 110 customer selectable gas combinations
- Bi-directional RS-232 serial digital port for use with a computer or other digital communication device
- 110 VAC or 220 VAC, 50/60 Hz (Specify at time of order)